AII-2000 M OXYGEN ANALYZER OR PSR-11-39-JD2

 

Portable percent oxygen analyzer measures O2 concentrations from 0.05% to 100%. Features an advanced galvanic oxygen sensor, a longer operating life, a true 18 month warranty in normal use, proven electronics with touch pad controls, an external sensor with a coiled cable that extends to 6 feet, and two user selectable alarm set points with a 15% lower limit.

 

Accuracy: < +1% of FS range under constant conditions  Analysis:  0-100% oxygen
 Application:  Anesthesia
Respiratory Therapy – Ventilators, Respirators
Neonatal & Pediatric Incubators & Hoods
Oxygen Therapy – Intensive Care
Emergency Transport
 Approvals:  ISO 9001:2000, MDD 93/42/Annex II, ISO 13485:2003
 Area Classification:  General purpose  Alarms:  User adjustable HI alarm 16-100% and LO alarm 15-99%; 120 second alarm silence for calibration purposes HI alarm defeat for flushing patients with 100% O2
 Calibration:  Air or certified 100% O2every 8 hours; time – 90 seconds  Compensation:  Temperature compensated
 Connections:  1×16 mm thread or push-in flow diverter with o-ring seal  Controls:  Soft touch keypad for ON/OFF and menu function
 Dimensions:  3.6 x 5.9 x 1.6”; weight 10 oz. (280 grams)  Display:  3-1/2 digit backlit LCD 2.75 x 1.375; resolution 0.1% O2
 Flow Sensitivity:  None between 0.2 to 10 liters per minute    
 LED Indicators:  None  Humidity:  Non-condensing 0-95% RH
 Linearity:  +1% under constant conditions  Pressure:  Inlet – ambient; vent – atmospheric
 Power:  2 AA Alkaline batteries; 1,200 hours continuous use  Response Time:  90% of final FS reading in 9 seconds
Sensitivity:  < 0.5% of FS range  Sensor:  AII-11-60
 Sensor Life:  60 months in air at 25ºC and 1 atmosphere  Signal Output:  None
 Storage Temp.:  -20º to 60ºC (-4ºF to 140ºF) on intermittent basis  Temp. Range:  5º to 45ºC (41ºF to 113ºF)
 Warm-up Time:  None  Warranty:  24 months analyzer; 18 months sensor (any application)