GPR-1100 INTRINSICALLY SAFE PORTABLE PPM O2 ANALYZER
ATEX certified portable PPM oxygen analyzer, 0-10 PPM low range,. features certification to ATEX Directive 94/9/EC. Features the proprietary micro-fuel cell type PPM oxygen sensor, field proven for reliably, a 24 month operating sensor life with no periodic maintenance, user friendly digital electronics, and a stainless steel 4-way sample/bypass system that enables the analyzer to come online at low PPM levels in minutes.
Certification:
Ex ia IIB T4 Gb
Tamb -10oC to 45oC
Accuracy: | +/- 2% of FS | Analysis Ranges: | 0-10 PPM, 0-100, 0-1000 PPM, 1% and 25% |
Sensitvity: | <0.5% of lowest full scale range | Linearity: | >0.995 over all ranges |
Application: | Oxygen analysis in inert, helium, hydrogen and mixed gas streams | Approvals: | Ex ia IIB T4 Gb, Tamb -10oC to 45oC |
Area Classification: | Class I, Division 1, Groups C, D | Calibration: | Certified span gas |
Compensation: | Barometric pressure and temperature | Connections: | 1/8″ compression tube fittings |
Controls: | Menu driven range selection, calibration and system functions | Display: | Graphical LCD 2.75 x 1.375”; resolution .01 PPM |
Enclosure: | Painted aluminum NEMA 4X, 4.6 x 9 x 3″, 10 lbs. | Sample Flow: | Recommended flow rate 1-2 SCFH |
LED Indicators: | LOW BATT (72 hr. warning); CHARGE mode | Signal Output: | 0-1 V DC |
Pressure: | Inlet – regulate to 5-30 psig; vent – atmosphere | Power: | Rechargeable battery, 60 day cycle |
Recovery Time: | 30 seconds in air to < 10 PPM in < 60 min on N2 purge | Response Time: | 90% of final FS reading in 10 seconds |
Sample System: | Quick disconnect fittings | Wetted Parts | Stainless steel |
Sensor Model: | GPR-12-333 | Sensor Life: | 24 months |
Temp Range: | 5oC to 45oC | Temp. Range: | 12 months analyzer, 12 months sensor |